Inaccuracy of seven popular sphygmomanometers for home measurement of blood pressure
- 30 June 1990
- journal article
- research article
- Published by Wolters Kluwer Health in Journal Of Hypertension
- Vol. 8 (7) , 621-634
- https://doi.org/10.1097/00004872-199007000-00005
Abstract
Seven models, available commercially for the self-measurement of blood pressure, were subjected to a validation procedure in which three devices of each model were tested by observers who were trained to a high standard of accuracy. The models were the Omron HEM-400C, the Philips HP5308, the Healthcheck ''Cuffless'' CX-5 060020, the Nissei Analogue Monitor, the Philips HP5306/B, the Systema Dr MI-150 and the Fortec Dr MI-100. The validation programme had a number of unique features which included assessment of interdevice variability before and after 1 month of home use, and a new form of analysis, which we term ''clinical'', based on the likely influence of three grades of device inaccuracy on patient management. In the main validation phase, one device of each model was compared with simultaneous measurements made by two ''blinded'' observers using a standard mercury sphygmomanometer (PyMaH Corporation, New Jersey, USA) in the same arm in 85 subjects with a wide range of blood pressures. Three models (the Healthcheck ''Cuffless'' CX-5 060020, the Systema Dr MI-150 and the Fortec Dr MI-100) failed the interdevice variability tests and did not reach the main validation test. Two models (the Omron HEM-400C and the Philips HP5306/B) failed on the criteria set down by the American National Standard for Electronic or Automated Sphygmomanometers, as well as the ''clinical'' criteria. The remaining two models (the Nissei Analogue Monitor and the Philips HP5308) were acceptable for the measurement of systolic blood pressure by both methods of analysis but failed in the ''clinical'' analysis for diastolic blood pressure. The mercury sphygmomanometer was comfortably within the criteria for both methods of analysis.This publication has 5 references indexed in Scilit:
- Home blood pressure-measuring devices: A comparative study of accuracyJournal Of Hypertension, 1989
- A comparison of the random-zero and standard mercury sphygmomanometers.Hypertension, 1988
- Prevalence and accuracy of home sphygmomanometers in an urban population.American Journal of Public Health, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Concurrent validity studies of blood pressure instrumentation. The Philadelphia Blood Pressure Project.Hypertension, 1984