Use of an oxygen concentrator in a Nigerian neonatal unit: economic implications and reliability
- 1 September 2002
- journal article
- Published by Taylor & Francis in Paediatrics and International Child Health
- Vol. 22 (3) , 209-212
- https://doi.org/10.1179/027249302125001499
Abstract
A 3-year experience of using an oxygen concentrator in a Nigerian newborn unit and economic appraisal of its effectiveness is reported. The oxygen concentrator is a device that absorbs nitrogen from ambient air, with a resultant oxygen concentration of 85 to 95% at different flow rates. The oxygen concentrator met our oxygen needs which averaged 18 hours a day, and had a huge cost advantage over the oxygen cylinders. The cost of oxygen via cylinder for just one patient for a year exceeds the initial capital outlay for a concentrator. The Puritan-Bennett oxygen concentrator has a lifespan of at least 7 years and is virtually maintenance-free for the 1st 26,400 hours of use, after which some major components might need replacement. We conclude that in developing countries oxygen concentrators are a more cost-effective, reliable and convenient means of oxygen supply than oxygen cylinders, and recommend their use where there is a high demand for oxygen.Keywords
This publication has 6 references indexed in Scilit:
- Resuscitation of newborn infants with room air or oxygenSeminars in Neonatology, 2001
- International Guidelines for Neonatal Resuscitation: An Excerpt From the Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care: International Consensus on SciencePediatrics, 2000
- ILCOR Advisory Statement: Resuscitation of the Newly Born InfantCirculation, 1999
- Oxygen Concentrator in a Remote Hospital in ZaïreTropical Doctor, 1985
- Oxygen ConcentratorsInternational Anesthesiology Clinics, 1982
- Problems with Equipment in Oxygen TherapyInternational Anesthesiology Clinics, 1982