System Requirements for High-Resolution Gamma-Ray Spectrometry at High Counting Rates
- 1 January 1969
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 16 (5) , 58-67
- https://doi.org/10.1109/TNS.1969.4325475
Abstract
The full potential of high-resolution gamma-ray spectrometers has been severely hampered by degradation in the quality of the pulse-height spectrum at high input data rates. Over the past year or two considerable effort has been expended in the examination of this problem. This paper will briefly review the nature of these instrumental difficulties, define realistic experimental requirements, and illustrate state-of-the-art solutions to these problems. A DC-coupled system which has been developed in our laboratory for high-rate applications will be described and performance presented.Keywords
This publication has 6 references indexed in Scilit:
- Charge Collection in Coaxial Ge(Li) DetectorsIEEE Transactions on Nuclear Science, 1968
- Optimum Signal-Processing for Pulse-Amplitude Spectrometry in the Presence of High-Rate Effects and NoiseIEEE Transactions on Nuclear Science, 1968
- Active D-C Restoration in Nuclear Pulse SpectrometryIEEE Transactions on Nuclear Science, 1967
- A High Precision DC RestorerIEEE Transactions on Nuclear Science, 1967
- A Low-Noise Charge Sensitive Preamplifier for Semiconductor Detectors Using Paralleled Field-Effect-TransistorsIEEE Transactions on Nuclear Science, 1966
- Reduction of Baseline Shift in Pulse-Amplitude MeasurementsReview of Scientific Instruments, 1961