Heat Sink Protection against Lasering Endotracheal Cuffs
- 1 November 1982
- journal article
- Published by SAGE Publications in Annals of Otology, Rhinology & Laryngology
- Vol. 91 (6) , 606-607
- https://doi.org/10.1177/000348948209100617
Abstract
The ignition of an endotracheal cuff by CO2 laser impaction has proved troublesome in several ignition sequences. Metal foil tape wrapped around the shaft of the tube will provide adequate protection for it, but the tape does not extend to the thin-walled cuff. When normal saline solution is used instead of air to fill the cuff, heat is rapidly conducted away from the wall and the cuff is prevented from reaching its kindling temperature. If penetration should occur, a fine jet of saline solution instantaneously sprays the field, extinguishing any early combustion. The reaction time of the surgeon or an assistant is not a factor in prevention of thermal injury, no new equipment is required and preparation time is minimal.Keywords
This publication has 9 references indexed in Scilit:
- Prevention of Fire Hazards Associated with Use of Carbon Dioxide LasersAnesthesia & Analgesia, 1981
- Laser-induced Endotracheal Tube FireAnesthesiology, 1981
- Incendiary Characteristics of Endotracheal Tubes with the Carbon Dioxide LaserAnnals of Otology, Rhinology & Laryngology, 1981
- Anaesthetic safety with the carbon dioxide laserAnaesthesia, 1981
- Complications of CO2 Laser Surgery of the LarynxAnnals of Otology, Rhinology & Laryngology, 1981
- Endotracheal Tube Ignition During Laser Surgery of the LarynxJAMA Otolaryngology–Head & Neck Surgery, 1979
- Ignition of a tracheal tube during laryngeal laser surgeryAnaesthesia, 1979
- Endotracheal Intubation and Venturi (Jet) Ventilation for Laser Microsurgery of the LarynxAnnals of Otology, Rhinology & Laryngology, 1976