Phase shifter calibration in phase-shifting interferometry
- 15 September 1985
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 24 (18) , 3049-3052
- https://doi.org/10.1364/ao.24.003049
Abstract
This paper describes some practical methods to calibrate the phase shifter in phase-shifting interferometry (PSI). The phase shifter used in the experiment is a piezoelectric transducer (PZT) that has a nonlinearity of σ value is 0.139°. A calibration-insensitive phase calculation algorithm is discussed and compared with other synchronous detection equations (e.g., the three-bucket or the four-bucket method). Experimental results verify the calibration-insensitive mechanism of the self-calibrating algorithm.Keywords
This publication has 2 references indexed in Scilit: