Immunonephelometry of apolipoprotein B with a centrifugal analyzer.
- 1 October 1987
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 33 (10) , 1882-1884
- https://doi.org/10.1093/clinchem/33.10.1882
Abstract
We developed an automated immunonephelometric assay for the measurement of apolipoprotein B (apo B) with a light-scattering microcentrifugal analyzer. Pretreating specimens with a dilute solution of Tween 20 or triglyceride lipase decreased the nephelometric response of apo B. Polyethylene glycol is included in the reaction mixture, and the reaction is complete within 4 min. The method is precise (CV = 6.5%, mean = 0.68 g/L) and the standard curve is linear to an apo B concentration of 2.8 g/L. Lipemia does not interfere with the method if grossly lipemic specimens are centrifuged to remove chylomicrons.This publication has 0 references indexed in Scilit: