THE EFFECT OF REPAIRS ON THE PERFORMANCE OF THE WRIGHT RESPIROMETER
Open Access
- 1 December 1970
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 42 (12) , 1127-1130
- https://doi.org/10.1093/bja/42.12.1127
Abstract
Respirometers are repaired in this department. They do not always retain the makers' specifications but they are clinically useful. A random selection of those in clinical use has shown that, on the average, the respirometers read 2 per cent low at low tidal volumes and 4 per cent high at high tidal volumes. Individual meters differed from the mean by up to ±15 per cent. Damage, caused by dropping, is responsible for the majority of technical failures.Keywords
This publication has 0 references indexed in Scilit: