Piezoelectric driven Kelvin probe for contact potential difference studies
- 1 July 1976
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 47 (7) , 840-842
- https://doi.org/10.1063/1.1134750
Abstract
A convenient piezoelectric driven Kelvin probe for the measurement of work function changes is described. The probe has a simple construction, is small, bakeable, and can be mounted on any 35‐mm‐i.d. UHV flange. As the piezoelectric device is inside the vacuum chamber, only one electrical feedthrough is needed for operating the probe. The distance between reference electrode and sample can be varied within ±0.5 mm simply by applying a dc voltage to the piezoelectric device in addition to the ac oscillator‐driving voltage, not requiring any mechanical driving and feedthrough provisions. The probe was tested studying the adsorption of O2 on Pt(111). The sensitivity for contact potential measurements is better than 0.1 mV at a response time of 10 sec.Keywords
This publication has 6 references indexed in Scilit:
- A new pendulum device to measure contact potential differencesApplied Physics A, 1974
- An integrated, high vacuum, beam modulation deviceJournal of Physics E: Scientific Instruments, 1973
- A critique of the Kelvin method of measuring work functionsJournal of Physics E: Scientific Instruments, 1970
- Stress Dependence of Contact Potential: The ac Kelvin MethodReview of Scientific Instruments, 1970
- A NEW METHOD OF MEASURING CONTACT POTENTIAL DIFFERENCES IN METALSReview of Scientific Instruments, 1932
- V. Contact electricity of metalsJournal of Computers in Education, 1898