Vibration-rotation structure in absorption bands for the calibration of spectrometers from 2 to 16 microns
- 1 January 1960
- journal article
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Bureau of Standards Section A: Physics and Chemistry
- Vol. 64A (1) , 29-48
- https://doi.org/10.6028/jres.064a.004
Abstract
Suitable bands of common gases have been tabulated and remeasured wherever necessary from 2 to 16 microns to obtain an accuracy of about 0.03 cm-1 throughout the region and to provide good calibrating points at frequent intervals. Some 600 rotation-vibration lines are illustrated in 20 spectrograms and wavenumbers are listed in companion tables with considerable intercomparison with worthy data obtained in other laboratories. The absorpotion bands were remeasured or calibrated by using either a precisely graduated grating circle or standard atomic lines with the fringe system formed by a Fabry-Perot interferometer. Characteristic features of the individual bands are discussed briefly and references to other publications are given. The substances used for calibration include H2O, CO2, CO, HCl, HBr, NH3, C2H2, CH4, N2O, and polystyrene film.Keywords
This publication has 1 reference indexed in Scilit:
- A Calibration Method for Infra-Red Prism Spectrometers1Journal of the Optical Society of America, 1948