Electromagnetic Shielding of High Voltage Laboratories
- 1 May 1971
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Power Apparatus and Systems
- Vol. PAS-90 (3) , 1400-1406
- https://doi.org/10.1109/tpas.1971.292944
Abstract
A modern high voltage laboratory has to be shielded mainly to avoid the electromagnetic disturbances of the partial discharge measurements. The paper describes the practical methods of shielding a high voltage laboratory and introduces a simplified method for shielding calculations which takes into account the effect of the seams, conducting gaskets, short-circuited slots and sound absorbing holes. A numerical example illustrates the results in the case of a large hall.Keywords