Physical characteristics of the ECAT 953B/31: a new high resolution brain positron tomograph
- 1 January 1991
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 10 (4) , 499-504
- https://doi.org/10.1109/42.108583
Abstract
A high spatial resolution brain positron tomograph, the ECAT 953B/31, is presented. The machine consists of two 76.5 cm diameter rings (patient port diameter: 36 cm) made of 24 8x8 BGO detector blocks each, each block being coupled to four 1 in(2) Hammamatsu phototubes. The machine has 15 9 cmx1 mm motorized removable septa, which allows reconstruction of 31 slices 3.375 mm apart. The transaxial resolution (FWHM) in the reconstructed image (with wobbling) is 4.5 mm at center, 5.2 mm (tangential) and 5.6 mm at 10 cm from center. With septa out of the field of view (FOV), the axial resolution degrades to 5.5 mm a center and 6.3 mm at 10 cm from center. At 1 muCi/cc with a 250 keV threshold and a 12 ns coincidence window, the sensitivity is 146100 true events, the ratio of randoms to trues is 0.10, the scatter fraction is 17% and the dead-time losses are 30%. With septa out of the FOV, the sensitivity is increased by a factor 3.6, while the scatter fraction reaches 0.9. Images obtained with (18 )F-DG and H(2)(15)O in human brains and (18 )F-DOPA in a baboon brain demonstrate that the ECAT 953B/31 will be suitable for high resolution and for low count rate brain studies.Keywords
This publication has 11 references indexed in Scilit:
- Physical characteristics of TTV03, a new high spatial resolution time-of-flight positron tomographIEEE Transactions on Nuclear Science, 1990
- Evaluation of the Karolinska new positron camera system; the Scanditronix PC2048-15BIEEE Transactions on Nuclear Science, 1990
- Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents countsIEEE Transactions on Nuclear Science, 1990
- Development of a high resolution PETIEEE Transactions on Nuclear Science, 1990
- Design of a super fast three-dimensional projection system for positron emission tomographyIEEE Transactions on Nuclear Science, 1990
- Design and evaluation of HEADTOME-IV, a whole-body positron emission tomographIEEE Transactions on Nuclear Science, 1989
- Performance measurements of the SP3000/UW time-of-flight positron emission tomographIEEE Transactions on Nuclear Science, 1988
- Physical performance of the latest generation of commercial positron scannerIEEE Transactions on Nuclear Science, 1988
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1986
- Correction for Scattered Radiation in a Ring Detector Positron Camera by Integral Transformation of the ProjectionsJournal of Computer Assisted Tomography, 1983