Abstract
Background— The US Food and Drug Administration recently issued a warning of subacute thrombosis and hypersensitivity reactions to sirolimus-eluting stents (Cypher). The cause and incidence of these events have not been determined. Methods and Results— We present findings of a 58-year-old man who died of late stent thrombosis 18 months after receiving 2 Cypher stents for unstable angina. Although angiographic and intravascular ultrasound results at 8 months demonstrated the absence of neointimal formation, vessel enlargement was present. An autopsy showed aneurysmal dilation of the stented arterial segments with a severe localized hypersensitivity reaction consisting predominantly of T lymphocytes and eosinophils. Conclusions— The known pharmacokinetic elution profile of Cypher stents and the presence of polymer fragments surrounded by giant cells and eosinophils suggest that a reaction to the polymer may have caused late stent thrombosis. Careful long-term follow-up of patients with vessel enlargement after Cypher stent placement is recommended.Keywords
This publication has 23 references indexed in Scilit:
- Long-Term Follow-Up of Incomplete Stent Apposition in Patients Who Received Sirolimus-Eluting Stent for De Novo Coronary LesionsCirculation, 2003
- Sirolimus-Eluting Stents versus Standard Stents in Patients with Stenosis in a Native Coronary ArteryNew England Journal of Medicine, 2003
- Regional Remodeling as the Cause of Late Stent MalappositionCirculation, 2003
- New Frontiers in CardiologyCirculation, 2003
- New Frontiers in CardiologyCirculation, 2003
- Pilot Trial of Oral Rapamycin for Recalcitrant RestenosisCirculation, 2003
- Sirolimus-induced leukocytoclastic vasculitisTransplantation, 2002
- Granulomatous ‘foreign body reactions’ contribute to exaggerated in-stent restenosisCoronary Artery Disease, 1999
- Coronary stent coatingsCoronary Artery Disease, 1994
- Rapamycin inhibits airway leukocyte infiltration and hyperreactivity in guinea pigsInflammation Research, 1993