Decay time measurement for high-Q ultrasonic resonators
- 1 March 1976
- journal article
- letter
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 47 (3) , 378-379
- https://doi.org/10.1063/1.1134641
Abstract
Decay time and half‐power bandwidth of high‐Q ultrasonic resonators can be determined with a calibrated RC network generating an exponential time base on a CRT.Keywords
This publication has 5 references indexed in Scilit:
- High Q ultrasonic liquid resonators with concave transducersReview of Scientific Instruments, 1976
- New acoustic resonator for liquids in the 0.2− to 2−MHz rangeThe Journal of the Acoustical Society of America, 1975
- Ultrasonic Measurements with Milliliter Liquid Samples in the 0.5–100 MHz RangeReview of Scientific Instruments, 1973
- Simple Method of Evaluating Exponential Relaxation Functions ElectronicallyNature Physical Science, 1971
- Ultrasonic reverberation measurements in liquidsApplied Scientific Research, Section B, 1950