Flow analysis method for the determination of silicic acid in highly purified water by gel-phase absorptiometry with molybdate and Malachite Green
- 1 January 1991
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 116 (8) , 835-840
- https://doi.org/10.1039/an9911600835
Abstract
A flow analysis method based on the on-line concentration and direct absorptiometric measurement of the ion-associate species formed between molybdosilicate and Malachite Green on Sephadex LH-20 gel, is described for the determination of silicic acid in highly purified water. The increase in the attenuance of the coloured species, which was concentrated on-line onto the gel packed in a flow-through cell, could be measured continuously with a spectrophotometer at 627 nm. The coloured species on the gel was easily desorbed with a mixture of sulphuric acid and acetone, and the gel-packed flow-through cell could be used repeatedly. The concentrations of silicic acid in 4–5 samples of highly purified water were determined in less than 1 h. The detection limit was 0.1 µg dm–3 of Si using water samples with a volume of about 5 cm3. The total amount of silicate-silicon in highly purified water was determined after treating the samples with ammonia solution.Keywords