Evaluation of a new VUV/soft x-ray toroidal grating monochromator with a movable exit slit
- 1 January 1992
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 63 (1) , 1269-1273
- https://doi.org/10.1063/1.1143097
Abstract
The performance of a toroidal grating monochromator can be significantly improved by compensating for the wavelength dependent defocusing aberration by means of a movable exit slit. Model calculations of a movable exit slit design show a resolving power that varies nearly linearly with wavelength over the scan range. The energy resolution follows the slope between two optimized resolution minima of a fixed slit TGM. The overall improvement in resolution due to the movable slit depends critically on the figure error of the toroid surface. The Al(LII,III) core edges provide a means to calibrate gratings and determine the resolution at 72.8 eV. A method is described for using the shift in core edge with exit slit position to enable final alignment of the exit slit motion providing stability of energy calibration of 2 meV over the scan range. Comparison of calculations and measured resolution suggests that slope errors remain the limiting factor in performance.Keywords
This publication has 15 references indexed in Scilit:
- A 6 m toroidal-grating-monochromator beam line for high momentum-resolution photoelectron spectroscopyNuclear Instruments and Methods in Physics Research, 1984
- Design and performance of a toroidal grazing incidence monochromator for the 20 – 200 eV photon energy rangeNuclear Instruments and Methods in Physics Research, 1982
- Design and performance of a beam line with a 1 m focal length toroidal grating monochromator at the ACO storage ringNuclear Instruments and Methods in Physics Research, 1982
- Design optimization of “straight groove” toroidal grating monochromators for synchrotron radiationNuclear Instruments and Methods, 1980
- Performance of a dual toroidal grating monochromatorNuclear Instruments and Methods, 1980
- Thermomodulation of the aluminum L2,3 absorption edgeSolid State Communications, 1979
- Design and Performance of Holographic Concave GratingsJapanese Journal of Applied Physics, 1976
- Metal Photocathodes as Secondary Standards for Absolute Intensity Measurements in the Vacuum Ultraviolet*Journal of the Optical Society of America, 1966
- The Torus GratingJournal of the Optical Society of America, 1950
- The Theory of the Concave GratingJournal of the Optical Society of America, 1945