Abstract
In diagnostic examinations of the trunk and head, the energy imparted to the patient is related to the radiation risk. In this work, the energy imparted to laterally infinite, 10-300 mm thick water slabs by 5-300 keV photons is calculated using a Monte Carlo photon transport model. The energy imparted is also derived for energy spectra of primary photons relevant to diagnostic radiology. In addition to values of energy imparted, values of backscattered and transmitted energies, quantities primarily obtained in the transport calculations, are reported. Assumptions about coherent scattering are shown to be important for values of backscattered and transmitted energies but unimportant with respect to values of energy imparted. Comparisons are made with other Monte Carlo results from the literature. Discrepancies of 10-20% in some calculated quantities can be traced back to the use of different tabulations of interaction cross-sections by various authors.