Pilot study of the accuracy of bedside glucometry in the intensive care unit
- 1 November 2001
- journal article
- research article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 29 (11) , 2205-2207
- https://doi.org/10.1097/00003246-200111000-00025
Abstract
To evaluate the accuracy of bedside glucometry among critically ill patients. Prospective audit, with sequential specimen collection. Thirty-two-bed cardiovascular, neurosurgical, and medical-surgical intensive care unit in a single Canadian center. Ten critically ill adults, who were sequentially followed during their intensive care unit stay. Eight had diabetes mellitus, and three were in shock. Repeat arterial blood samples were obtained by the attending nurse, who withdrew a portion of the sample and performed reflectance glucometry at the bedside. The remainder was immediately sent in a vacuum-sealed plasma separation tube to the hospital laboratory, and analyzed using a conventional plasma glucose analyzer by a laboratory technologist. Sequential samples were taken at intervals of at least 12 hrs of one other. A total of 105 arterial glucose pairs were obtained. There was a significant correlation between the laboratory and glucometry determined glucose concentrations (intraclass correlation coefficient = 0.86, p r Bedside glucose testing of arterial whole blood samples may be an accurate alternative to laboratory plasma glucose measurement among critically ill adults, within approximately 2.3 mmol/L of certainty. Because previous studies have suggested that this bedside technique may be prone to a moderate degree of error among patients in shock as well as those with an abnormal blood pH or hematocrit, larger studies are needed to confirm our findings.Keywords
This publication has 12 references indexed in Scilit:
- Effects of pH on Glucose Measurements With Handheld Glucose Meters and a Portable Glucose Analyzer for Point-of-Care TestingArchives of Pathology & Laboratory Medicine, 2000
- Stress hyperglycaemia and increased risk of death after myocardial infarction in patients with and without diabetes: a systematic overviewThe Lancet, 2000
- Point-of-Care Glucose TestingArchives of Pathology & Laboratory Medicine, 2000
- Clinical and statistical evaluation of self-monitoring blood glucose meters.Diabetes Care, 1998
- Prospective randomised study of intensive insulin treatment on long term survival after acute myocardial infarction in patients with diabetes mellitusBMJ, 1997
- Evaluating laboratory usage in the intensive care unitCritical Care Medicine, 1997
- A Comparison of Four Glucose Monitors in a Hospital Medical Surgical SettingClinical Nurse Specialist, 1993
- Fingerstick Glucose Determination in ShockAnnals of Internal Medicine, 1991
- Relationship Between Admission Hyperglycemia and Neurologic Outcome of Severely Brain-Injured PatientsAnnals of Surgery, 1989
- Intraclass correlations: Uses in assessing rater reliability.Psychological Bulletin, 1979