The Design of Filters to Produce “Flat” X-ray Isodose Curves at a Given Depth
- 1 December 1945
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 18 (216) , 382-385
- https://doi.org/10.1259/0007-1285-18-216-382
Abstract
A method of producing “flat” X-ray isodose curves at any given depth in a phantom by means of “stepped” copper filters, is described. Details of these filters as fitted in a variety of applicators are given. A specimen isodose curve is shown and the effects of the filter on the surface dose and dose distribution, and on the central percentage depth dose values, are discussed.Keywords