Continuous Electrophoretic Purification of Individual Analytes from Multicomponent Mixtures
- 9 March 2004
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 76 (8) , 2298-2305
- https://doi.org/10.1021/ac0350460
Abstract
Individual analytes can be isolated from multicomponent mixtures and collected in the outlet vial by carrying out electrophoretic purification through a capillary column. Desired analytes are allowed to migrate continuously through the column under the electric field while undesired analytes are confined to the inlet vial by application of a hydrodynamic counter pressure. Using pressure ramping and buffer replenishment techniques, 18% of the total amount present in a bulk sample can be purified when the resolution to the adjacent peak is approximately 3. With a higher resolution, the yield could be further improved. Additionally, by periodically introducing fresh buffer into the sample, changes in pH and conductivity can be mediated, allowing higher purity (>or=99.5%) to be preserved in the collected fractions. With an additional reversed cycle of flow counterbalanced capillary electrophoresis, any individual component in a sample mixture can be purified providing it can be separated in an electrophoresis system.Keywords
This publication has 25 references indexed in Scilit:
- A quantitative study of continuous flow‐counterbalanced capillary electrophoresis for sample purificationElectrophoresis, 2003
- Continuous free flow electrophoresis for preparative chiral separations of piperoxan using sulfated β-cyclodextrinThe Analyst, 2000
- Fraction collection in micropreparative capillary zone electrophoresis and capillary isoelectric focusingElectrophoresis, 2000
- Application of free‐flow electrophoresis for isolation and purification of proteins and peptidesElectrophoresis, 1998
- Micropreparative capillary electrophoresis of DNA by direct transfer onto a membraneElectrophoresis, 1997
- Micropreparation of peptides by capillary electrophoresis for matrix assisted laser desorption mass spectrometryFEBS Letters, 1996
- Design of a High-Precision Fraction Collector for Capillary ElectrophoresisAnalytical Chemistry, 1995
- Sodium chloride in separation medium enhances cell compatibility of free flow electrophoresisElectrophoresis, 1995
- Flow Counterbalanced Capillary ElectrophoresisAnalytical Chemistry, 1994
- Interaction of capillary zone electrophoresis with a sheath flow cuvette detectorAnalytical Chemistry, 1990