High survival rate of hydroxyapatite-coated hip prostheses100 consecutive hips followed for 10 years
- 1 January 2004
- journal article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 75 (2) , 127-133
- https://doi.org/10.1080/00016470412331294365
Abstract
We followed 100 consecutive primary total hip replacements with a proximal hydroxyapatite coating for 10 years. No patient was lost to follow-up. 29 patients (32 hips) died before the 10-year follow-up was done; none of their hips had been revised. Thigh pain on activity occurred in 3 hips. We found no radiographic signs of loosening of the femoral components. In course of time the location of dense bone around the femoral stem, which would suggest implant/bone stress transfer, moved distally in 51 of 67 stems after 10 years. No linear or distal osteolysis occurred around the stem. Revision of 1 stem was performed because of thigh pain, but it was found fixed to bone proximally, while 3 cups were revised because of acetabular osteolysis. The 10-year survival of the stem and cup, using revision or pending revision as endpoint, was 100% (95% CI: 99-100) and 97% (95% CI: 94-99), respectively.Keywords
This publication has 18 references indexed in Scilit:
- Hydroxyapatite-Coated Acetabular ComponentsJournal of Bone and Joint Surgery, 2001
- Survival analysis of cementless grit-blasted titanium total hip arthroplastiesThe Journal of Bone and Joint Surgery. British volume, 2001
- Total hip arthroplasty with an uncemented femoral component: Excellent results at 10-year follow-upThe Journal of Bone and Joint Surgery. British volume, 1997
- Osteogenesis imperfecta – where next?The Journal of Bone and Joint Surgery. British volume, 1997
- Hydroxyapatite and Fluorapatite Coatings for Fixation of Weight Loaded ImplantsClinical Orthopaedics and Related Research, 1997
- The fully hydroxyapatite-coated total hip implant: Clinical and roentgenographic resultsThe Journal of Arthroplasty, 1996
- Remodeling of Bone around Hydroxyapatite-Coated Femoral Stems*Journal of Bone and Joint Surgery, 1996
- A Tapered Titanium Femoral Stem Inserted without Cement in a Total Hip Arthroplasty. Radiographic Evaluation and Stability*Journal of Bone and Joint Surgery, 1996
- The Effect of Stem Fit on Bone Hypertrophy and Pain Relief in Cementless Total Hip ArthroplastyPublished by Wolters Kluwer Health ,1989
- Ectopic Ossification Following Total Hip ReplacementJournal of Bone and Joint Surgery, 1973