The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry
Top Cited Papers
Open Access
- 1 July 2005
- journal article
- research article
- Published by BMJ in British Journal of Ophthalmology
- Vol. 89 (7) , 851-854
- https://doi.org/10.1136/bjo.2004.056622
Abstract
Aims: To evaluate the influence of central corneal thickness (CCT) on intraocular pressure (IOP) measurements made with the Goldmann applanation tonometer (GAT), Tono-Pen XL, ocular blood flow tonograph (OBF), and Canon TX-10 non-contact tonometer (NCT). Methods: CCT was recorded for either eye (randomly selected) of each of 105 untreated patients with ocular hypertension and glaucoma attending the glaucoma research unit at Moorfields Eye Hospital. For each of the selected eyes, IOP was measured with the GAT (two observers), Tono-Pen, OBF, and NCT in a randomised order. The relation of measured IOP and of inter-tonometer differences with CCT and subject age was explored by linear regression analysis. Results: A significant association between measured IOP and CCT was found with each instrument. The change in measured IOP for a 10 μm increase in CCT was 0.28, 0.31, 0.38, and 0.46 for the GAT, Tono-Pen, OBF, and NCT, respectively (all p⩽0.05). There was a significant association between the NCT/GAT differences and CCT, with a tendency of NCT to overestimate GAT in eyes with thicker corneas. There was a significant association between GAT/Tono-Pen and OBF/Tono-Pen differences and age, with a tendency of GAT and OBF to overestimate the Tono-Pen in eyes of older subjects. Conclusion: IOP measurement by all four methods is affected by CCT. The NCT is affected by CCT significantly more than the GAT. Subject age has a differential effect on the IOP measurements made by the GAT and OBF compared to the Tono-Pen.Keywords
This publication has 38 references indexed in Scilit:
- Varying Effects of Corneal Thickness on Intraocular Pressure Measurements with Different TonometersEye, 2004
- Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometerBritish Journal of Ophthalmology, 2004
- Central corneal thickness, radius of the corneal curvature and intraocular pressure in normal subjects using non‐contact techniques: Reykjavik Eye StudyActa Ophthalmologica Scandinavica, 2002
- Primary orbital melanoma masquerading as vascular anomaliesEye, 2002
- Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readingsBritish Journal of Ophthalmology, 2001
- The Influence of Corneal Thickness and Curvature on the Difference Between Intraocular Pressure Measurements Obtained with a Non-contact Tonometer and Those with a Goldmann Applanation TonometerJapanese Journal of Ophthalmology, 2000
- Tono-Pen tonometer and corneal thicknessEye, 1999
- Central corneal thickness and intraocular pressure in a Mongolian population,Ophthalmology, 1998
- Ageing of the human corneal stroma: structural and biochemical changesBiochimica et Biophysica Acta (BBA) - Molecular Basis of Disease, 1992
- Automatic Tonometer with Exact Theory: Various Biological ApplicationsScience, 1960