Comparison of two automated instrument systems for rapid susceptibility testing of gram-negative bacilli
- 1 December 1983
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 18 (6) , 1301-1309
- https://doi.org/10.1128/jcm.18.6.1301-1309.1983
Abstract
The Vitek AutoMicrobic System with GSC-plus cards and the Abbott MS-2 system were tested in parallel and the results were compared directly with those of a reference microdilution minimal inhibitory concentration (MIC) procedure on a group of 262 clinical isolates of the family Enterobacteriaceae and of Pseudomonas aeruginosa. Results of both systems were compared with the reference MIC for category agreement, and in addition, the Vitek MICs were compared with those obtained by the reference procedure. The Vitek system provided an essential category correlation of 89.4% for enteric bacteria and 97.0% for P. aeruginosa. Vitek MICs agreed within 1 twofold dilutional increment for 86.3% of the enteric bacteria tested and for 96.2% of the P. aeruginosa isolates. The Abbott MS-2 essential categoric agreement was 92.0% for enteric bacteria and 92.4% for P. aeruginosa. If only aminoglycosides or carbenicillin were considered for P. aeruginosa isolates, the essential category agreement was 92.5% for the Vitek and 93.3% for the MS-2. The majority of MS-2 category errors (13 of 19) with P. aeruginosa involved gentamicin results on isolates whose reference MICs were 8 micrograms/ml and whose MS-2 results were susceptible (MIC less than or equal to 4 micrograms/ml). Retesting of the P. aeruginosa isolates in calcium-supplemented MS-2 broth increased the essential agreement for the aminoglycosides to 97.5%.This publication has 16 references indexed in Scilit:
- Antimicrobial susceptibility testing (AST): A review of changing trends, quality control guidelines, test accuracy, and recommendation for the testing of β-lactam drugsDiagnostic Microbiology and Infectious Disease, 1983
- Comparison of three automated systems for antimicrobial susceptibility testing of gram-negative bacilliJournal of Clinical Microbiology, 1982
- Quantitative susceptibility test methods in major United States medical centersAntimicrobial Agents and Chemotherapy, 1981
- Clinical evaluation of automated antibiotic susceptibility testing with the MS-2 systemJournal of Clinical Microbiology, 1980
- Clinical laboratory evaluation of the Abbott MS-2 automated antimicrobial susceptibility testing system: report of a collaborative studyJournal of Clinical Microbiology, 1980
- Automated Microbiological Detection/Identification SystemJournal of Clinical Microbiology, 1977
- Relationship of Early Readings of Minimal Inhibitory Concentrations to the Results of Overnight TestsAntimicrobial Agents and Chemotherapy, 1975
- Laboratory Evaluation of a Rapid, Automated Susceptibility Testing System: Report of a Collaborative StudyAntimicrobial Agents and Chemotherapy, 1975
- Antibiotic susceptibility testing by a standardized single disk method.1966
- ANTIBIOTIC SUSCEPTIBILITY TESTING BY A STANDARDIZED SINGLE DISK METHOD1966