Limitation of Correction of Frequency Dependent Artefact in Pressure Recordings Using Harmonic Analysis
- 1 November 1974
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 50 (5) , 992-997
- https://doi.org/10.1161/01.cir.50.5.992
Abstract
Frequency dependent phase and amplitude distortion may theoretically be minimized using correction formulae derived from a model which assumes that water-filled catheter-manometer systems are analogous to an oscillating mass attached to a spring and damping pot. This study defines the practical limits of this correction technique using harmonic analysis of simultaneously recorded water-filled and catheter-tip manometer systems. Some improvement was obtained in tracings recorded with underdamped, high frequency response systems. In no case were exact duplications of the catheter-tip tracings obtained. Divergence of measured vitro frequency response from actual in vivo frequency response was one source of these discrepancies. Loose coupling of the components of the system also changed the behavior from that of the assumed simple mechanical model and at times prevented adequate correction. Catheter-tip manometers, in spite of their expense and relative fragility, still appear to be the best means at present of obtaining accurate pressure tracings, particularly at high heart rates or where time derivatives are to be computed.Keywords
This publication has 8 references indexed in Scilit:
- Damped and undamped frequency responses of underdamped catheter manometer systemsAmerican Heart Journal, 1970
- Evaluation and correction of manometer systems with two degrees of freedom.Journal of Applied Physiology, 1969
- AN AIR FILLED CATHETER FOR BIOOD PRESSURE MEASUREMENTThe Japanese Journal of Physiology, 1969
- Use of Fourier Series for the Analysis of Biological SystemsBiophysical Journal, 1966
- Harmonic analysis of pressure pulses obtained from the heart and great vessels of manAmerican Heart Journal, 1965
- THE PRESSURE WAVE FORM IN THE PULMONARY ARTERYHeart, 1963
- Evaluation of the pressure time derivative method for estimating peak blood flowAmerican Heart Journal, 1962
- Physiologic Recording by Modern Instruments With Particular Reference to Pressure RecordingPhysiological Reviews, 1960