Imaging Properties of a Positron Tomograph with 280 Bgo Crystals
Open Access
- 1 February 1981
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 28 (1) , 81-89
- https://doi.org/10.1109/tns.1981.4331144
Abstract
The basic imaging properties of the Donner 280-BGO-Crystal positron tomograph were measured and compared with the same system when it was equipped with 280 NaI(T1) crystals. The NaI(T1) crystals were 8 mm × 30 mm × 50 mm deep, sealed in 10 mm wide stainless steel cans. The BGO crystals are 9.5 mm wide × 32 mm × 32 mm deep and as they are not hygroscopic do not require sealed cans. With a shielding gap of 3 cm (section thickness 1.7 cm FWHM) the sensitivity of the BGO system is 55,000 events per sec for 1 μCi per cm3 in a 20 cm cylinder of water, which is 2.3 times higher than the NaI(T1) system. For a 200 μCi/cm line source on the ring axis in a 20 cm diameter water cylinder, the BGO system records 86% of the scatter fraction and 66% of the accidental fraction of the NaI(T1) system. The lower light yield and poorer time resolution of BGO requires a wider coincidence timing window than NaI(T1); however, the ability to use full-energy pulse height selection with a 2.3-fold improvement in sensitivity results in an overall reduction in the fraction of accidental events recorded. The in-plane resolution of the BGO system is 9-10 mm FWHM within the central 30 cm diameter field, and the radial elongation at the edge of the field in the NaI(T1) system has been nearly eliminated.Keywords
This publication has 16 references indexed in Scilit:
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1979
- Data Collection for Cross-Sectional Image Reconstruction by a Moving Ring of Positron Annihilation DetectorsJournal of Computer Assisted Tomography, 1979
- Trends and Prospects for Circular Ring Positron CamerasIEEE Transactions on Nuclear Science, 1979
- Sampling Requirements and Detector Motion for Positron Emission TomographyIEEE Transactions on Nuclear Science, 1979
- "Positology"-the search for suitable detector arrangements for a positron ECT with continuous rotationIEEE Transactions on Nuclear Science, 1979
- Engineering Aspects of PETT VIEEE Transactions on Nuclear Science, 1979
- The Donner 280-Crystal High Resolution Positron TomographIEEE Transactions on Nuclear Science, 1979
- Evaluation of a high resolution ring detector positron camera systemInternational Journal of Nuclear Medicine and Biology, 1978
- High Resolution Computed Tomography of Positron EmittersIEEE Transactions on Nuclear Science, 1977
- Three-dimensional Reconstruction from Radiographs and Electron Micrographs: Application of Convolutions instead of Fourier TransformsProceedings of the National Academy of Sciences, 1971