Abstract
A simple technique for measuring thermal conductivity of biomaterials is described. The method is based on depositing a pulse of heat into the material of choice, and fitting the subsequent local temperature decay to that predicted by a theoretical model. This transient method is most suitable in situations where frequent measurements of the thermal conductivity are desired. The method was evaluated by calculating the thermal conductivity of several inert materials. The measured conductivities compared well with published values. The developed technique was also used to examine the applicability of the “apparent conductivity” index to combine both conductive and blood-convective thermal effects in living, blood perfused tissues. Using both simulated and experimental results, it was shown that the changes in the apparent conductivity are highly correlated with changes in blood flow. However, quantitative application of this index must be restricted to conditions that are similar to those which existed at the time the apparent conductivity was measured.