An Evaluation of Long‐term Stimulation Thresholds by Measurement of Chronic Strength Duration Curves
- 1 July 1981
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 4 (4) , 376-379
- https://doi.org/10.1111/j.1540-8159.1981.tb03715.x
Abstract
Successful long-term cardiac pacemaker therapy requires both reliable generator output and stable, attainable chronic stimulation thresholds. Strength duration curves, which display cardiac stimulation thresholds as a function of impulse duration, provide needed data to evaluate more thoroughly acute and chronic lead thresholds. We measured chronic strength duration curves at the time of generator replacement in 47 patients with three different types of lead systems of variable geometry and surface area and compared these curves with acute strength duration curves measured at the time of initial implantation in 19 patients. We found that strength duration curves has a characteristic shape; however, chronic strength duration curves were shifted upward and to the right of acute strength duration curves, regardless of electrode tip surface area or geometric design. Our data suggest that variable pulse width exceeding 1 ms may not significantly reduce stimulation thresholds.Keywords
This publication has 3 references indexed in Scilit:
- Threshold behavior of electrodes in long-term ventricular pacingThe American Journal of Cardiology, 1977
- Cardiac pacing and pacemakers IV. Threshold of cardiac stimulationAmerican Heart Journal, 1977
- Clinical thresholds of endocardial cardiac stimulation: A long-term studyJournal of Surgical Research, 1975