Determination of Cadmium in Blood and Urine by Graphite Furnace Atomic Absorption Spectrophotometry

Abstract
We describe a method for measuring quantities of cadmium in blood, plasma, and urine by using graphite furnace atomic absorption preceded by wet ashing. The method appears to be extremely accurate and reproducible (coefficient of variation is 7% for 2.8 µg/liter concentration), with a detection limit of 2 pg.