Fully 3D Bayesian image reconstruction for the ECAT EXACT HR+
- 1 June 1998
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 45 (3) , 1096-1103
- https://doi.org/10.1109/23.681985
Abstract
[[abstract]]©1998 IEEE - A fully 3D Bayesian method is described for high resolution reconstruction of images from the Siemens/CTI ECAT EXACT HR+ whole body positron emission tomography (PET) scanner. To maximize resolution recovery from the system we model depth dependent geometric efficiency, intrinsic detector efficiency, photon pair non-colinearity, crystal penetration and inter-crystal scatter. We also explicitly model the effects of axial rebinning and angular mashing on the detection probability or system matrix. By fully exploiting sinogram symmetries and using a factored system matrix and automated indexing schemes, we are able to achieve substantial savings in both the storage size and time required to compute forward and backward projections. Reconstruction times are further reduced using multi-threaded programming on a four processor Unix server. Bayesian reconstructions are computed using a Huber prior and a shifted-Poisson likelihood model that accounts for the effects of randoms subtraction and scatter. Reconstructions of phantom data show that the 3D Bayesian method can achieve improved FWHM resolution and contrast recovery ratios at matched background noise levels compared to both the 3D reprojection method and an OSEM method based on the shifted-Poisson model.[[fileno]]2060214010001[[department]]醫環This publication has 14 references indexed in Scilit:
- Fast iterative image reconstruction of 3D PET dataPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2002
- Accurate geometric and physical response modelling for statistical image reconstruction in high resolution PETPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2002
- Evaluation of the ECAT EXCAT HR with ACSII for clinical routine 3D measurementsPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2002
- Performance evaluation of the whole-body PET scanner ECAT EXACT HR/sup +/ following the IEC standardIEEE Transactions on Nuclear Science, 1997
- Evaluation of simulation-based scatter correction for 3-D PET cardiac imagingIEEE Transactions on Nuclear Science, 1997
- Bayesian reconstruction of PET images: methodology and performance analysisPhysics in Medicine & Biology, 1996
- A system for the 3D reconstruction of retracted-septa PET data using the EM algorithmIEEE Transactions on Nuclear Science, 1995
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994
- Fast gradient-based methods for Bayesian reconstruction of transmission and emission PET imagesIEEE Transactions on Medical Imaging, 1994
- Parallelization of the EM algorithm for 3-D PET image reconstructionIEEE Transactions on Medical Imaging, 1991