Application of a Self-Tonometer to Home Tonometry
- 1 January 1986
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Ophthalmology (1950)
- Vol. 104 (1) , 49-53
- https://doi.org/10.1001/archopht.1986.01050130059020
Abstract
• Although home tonometry has been advocated in the management and diagnosis of glaucoma, the lack of proper instrumentation has precluded its routine performance. We have developed a self-tonometer that can be used by the patient at home and at work and have evaluated the performance of the instrument under these conditions. The results show that reproducibility of self-tonometer measurements is ± 1.4 mm Hg and is independent of the user. When the readings of the self-tonometer in a given individual were compared with those of the Goldmann tonometer, they varied by ±2.4 mm Hg. Patients who could be trained in the clinic performed well and safely on their own. An average of fewer than five attempts was necessary to obtain four readings at each time of day. Our preliminary findings obtained during home tonometry demonstrated some of the benefits of repeated diurnal curves.This publication has 13 references indexed in Scilit:
- Complex approach in early diagnosis of primary open angle glaucoma (clinical case)Practical medicine, 2018
- Evaluation of a Self Tonometer for Home UseArchives of Ophthalmology (1950), 1983
- The Effect of Hospitalization on Intraocular PressureAmerican Journal of Ophthalmology, 1982
- An Instrument for Self-Measurement of Intraocular PressureIEEE Transactions on Biomedical Engineering, 1982
- Long-Term Follow-Up of Treated Open-Angle GlaucomaAmerican Journal of Ophthalmology, 1979
- Clinical and statistical aspects on standardization of tonometersAlbrecht von Graefes Archiv für Ophthalmologie, 1979
- THE ACCURACY OF CLINICAL APPLANATION TONOMETRYActa Ophthalmologica, 1978
- OCULAR HYPERTENSION AND GLAUCOMATOUS OPTIC NERVE DAMAGEActa Ophthalmologica, 1974
- Home TonometryAmerican Journal of Ophthalmology, 1973
- The Significance of the Diurnal Tension Variations in Normal and Glaucomatous EyesArchives of Ophthalmology (1950), 1960