T2 Accuracy on a whole‐body imager
- 1 November 1997
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 38 (5) , 759-768
- https://doi.org/10.1002/mrm.1910380512
Abstract
MR oximetry requires a T2 measurement that is accurate within 5% in vivo. Simple methods are susceptible to signal loss and tend to underestimate T2. Current methods utilize RF pulses or RF cycling patterns that prevent signal loss at each data acquisition. However, using these methods with imperfect pulses, T2 tends to be overestimated due to temporary storage of the magnetization along the longitudinal axis where it decays more slowly with a time constant T1T2. To reduce the T1 dependence while preventing signal loss, we utilize simple 90x180y90x composite pulses and good RF cycling patterns. These trains are critical for T2 accuracy over typical ranges of RF and static field inhomogeneities and refocusing intervals. T1 signal decay during each 90x180y90x pulse must be accounted for to yield accuracy within 5% when the pulse‐width is 10% or more of the refocusing interval. A simple correction scheme compensates for this T1‐related error effectively.Keywords
This publication has 27 references indexed in Scilit:
- Coronary Angiography with Magnetization‐Prepared T2 ContrastMagnetic Resonance in Medicine, 1995
- In vivo visualization of myelin water in brain by magnetic resonanceMagnetic Resonance in Medicine, 1994
- 180° refocusing pulses which are insensitive to static and radiofrequency field inhomogeneityJournal of Magnetic Resonance (1969), 1992
- Estimating oxygen saturation of blood in vivo with MR imaging at 1.5 TJournal of Magnetic Resonance Imaging, 1991
- New, compensated Carr-Purcell sequencesJournal of Magnetic Resonance (1969), 1990
- Variation in measured transverse relaxation in tissue resulting from spin locking with the CPMG sequenceJournal of Magnetic Resonance (1969), 1988
- In vivo determination of T1 and T2 in the brain of patients with severe but stable multiple sclerosisMagnetic Resonance in Medicine, 1988
- Iterative carr-purcell trainsJournal of Magnetic Resonance (1969), 1988
- Compensation for pulse imperfections in NMR spin-echo experimentsJournal of Magnetic Resonance (1969), 1981
- Errors in measurements of transverse relaxation ratesJournal of Magnetic Resonance (1969), 1973