Errors in the measurements of T2 using multiple‐echo MRI techniques. II. Effects of static field inhomogeneity
- 1 August 1986
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 3 (4) , 562-574
- https://doi.org/10.1002/mrm.1910030410
Abstract
The accurate estimation of the spin‐spin relaxation time T2 is an important goal in magnetic resonance imaging particularly because it can be used for quantitative tissue characterization. The spin‐spin relaxation time T2 may be estimated using multiecho pulse sequences, but the accuracy of the estimate is dependent on the fidelity of the spin‐echo amplitudes, which may be severely compromised by rf pulse and static field imper‐fections. In this paper, the effects of static field inhomogeneities are investigated. The propagation of the errors introduced by off‐resonance effects are analyzed through computer simulations and analytical solutions of the Bloch equations. A series of experiments performed on a simple tissue phantom using a whole‐body imaging system operating at 6.35 MHz corroborates the simulation and analytical results. For accurate measurements of T2 using a whole‐body imaging system it is necessary to correct for these inhomogeneities. A correction scheme which would enable a more accurate estimate of T2 is currently under investigation. © 1986 Academic Press, Inc.This publication has 4 references indexed in Scilit:
- Errors in the measurements of T2 using multiple‐echo MRI techniques. I. Effects of radiofrequency pulse imperfectionsMagnetic Resonance in Medicine, 1986
- Symmetrical composite pulse sequences for NMR population inversion. II. Compensation of resonance offsetJournal of Magnetic Resonance (1969), 1982
- Symmetrical composite pulse sequences for NMR population inversion. I. Compensation of radiofrequency field inhomogeneityJournal of Magnetic Resonance (1969), 1982
- Compensation for pulse imperfections in NMR spin-echo experimentsJournal of Magnetic Resonance (1969), 1981