ASSESSMENT OF RADIATION EXPOSURE CAUSED BY TRANSMISSION SCANS IN SPECT: AN ANTHROPOMORPHIC DOSIMETRY STUDY
- 1 January 2002
- journal article
- abstracts
- Published by Walter de Gruyter GmbH in Biomedizinische Technik/Biomedical Engineering
- Vol. 47 (s1b) , 849-850
- https://doi.org/10.1515/bmte.2002.47.s1b.849
Abstract
Transmission measurements are performed in SPECT to correct for attenuation of the gamma quanta in the body. In this study, we measured the additional radiation dose caused by transmission scans using a field of collimated 153Gd rod sources. Two measurement series were performed with an anthropomorphic phantom and thermoluminescence dosimeters. For a typical SPECT study, we found a mean dose rate of 2.2 +/- 0.8 micro Sievert per hour (range: 1.2-3.9 micro Sievert per hour). For a measurement time of 20 min, this corresponds to a mean equivalent patient radiation dose of 0.73 micro Sievert. Thus, the radiation exposure caused by transmission scans can be neglected compared to the radiopharmaceutical dose and may not be considered as a limiting factor for the clinical application of attenuation correction in SPECT.Keywords
This publication has 4 references indexed in Scilit:
- An iterative transmission algorithm incorporating cross‐talk correction for SPECTMedical Physics, 2002
- Attenuation correction and gating synergistically improve the diagnostic accuracy of myocardial perfusion SPECTJournal of Nuclear Cardiology, 2002
- Dosimetry of transmission measurements in nuclear medicine: a study using anthropomorphic phantoms and thermoluminescent dosimetersEuropean Journal of Nuclear Medicine and Molecular Imaging, 1998
- Post-injection attenuation correction for PET scanners without measured random correctionIEEE Transactions on Nuclear Science, 1998