A scanning system for chest radiography with regional exposure control: Theoretical considerations
- 1 September 1983
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 10 (5) , 646-654
- https://doi.org/10.1118/1.595327
Abstract
Conventional chest radiography is limited by the presence of scattered radiation and the small useful exposure range of radiographic film. A computer-assisted scanning system to minimize these 2 effects is outlined. The system uses a small beam of radiation swept over the patient''s chest in a raster pattern to expose a conventional film cassette, while a slit collimator scanning between the patient and the film serves to reject scattered photons. A microcomputer measures beam attenuation by the patient with a detector placed behind the film which in turn automatically adjusts the x-ray tube output to minimize excursions in film exposure as the beam scans. A formalism which relates the patients transmission and film exposure distribution is developed and a system transfer function is given. Evidently, such a system operates as a spatial filter which attenuates film contrast for structures of spatial frequency less than the inverse scanning beam width. By manipulating the software parameters of the feedback network, it is possible to alter this filter and produce radiographs with low spatial frequency enhancement, attenuation or contrast inversion.This publication has 9 references indexed in Scilit:
- A scanning system for chest radiography with regional exposure control: Practical implementationMedical Physics, 1983
- Tantalum air-interspace crossed grid: design and performance characteristics.Radiology, 1982
- Generalized image combinations in dual KVP digital radiographyMedical Physics, 1981
- Contrast-detail-Dose and dose efficiency analysis of a scanning digital and a screen-film-grid radiographic systemMedical Physics, 1981
- Radiographic scanning in barium examinations of the gastrointestinal tract.Radiology, 1981
- Computerized fluoroscopy: digital subtraction for intravenous angiocardiography and arteriographyAmerican Journal of Roentgenology, 1980
- The Intensity of Scattered Radiation in MammographyRadiology, 1978
- A Method to Absorb Scattered Radiation Without Attenuation of the Primary BeamRadiology, 1976
- IDENTIFICATION OF POSITIVES IN ROENTGENOGRAPHIC READINGSPublished by Elsevier ,1960