The effective attenuation coefficient of soft tissue in the presence of Compton scattering from bone: experiments on models
- 1 July 1980
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 25 (4) , 711-717
- https://doi.org/10.1088/0031-9155/25/4/009
Abstract
Calibration methods are investigated for the determination of bone density by the Compton scattering method. The scattered radiation from materials of different densities and dimensions is measured for this purpose. A function for the dependance of the intensity on these parameters is derived and compared with the experimental results. The influence of the surrounding soft tissue is studied by simulation experiments with water. It is found that the 'effective attenuation coefficient' of water for the incident and scattered radiation depends on the thickness of the surrounding water layer itself, as well as on the density of the scattering material. The implications of these findings for the evaluation of bone density measurements are discussed.This publication has 8 references indexed in Scilit:
- Compton scatter densitometry in cancellous bonePhysics in Medicine & Biology, 1979
- Compton-scatter tissue densitometry: calculation of single and multiple scatter photon fluencesPhysics in Medicine & Biology, 1978
- The early detection of osteoporosis by Compton gamma ray spectroscopyPhysics in Medicine & Biology, 1977
- Bone density measured by photon scattering. II. Inherent sources of errorPhysics in Medicine & Biology, 1976
- Multiple scattering in the Compton effect. II. Analytic and numerical treatment of energy profilesPhysical Review A, 1976
- Multiple Scattering in the Compton EffectPhysical Review B, 1930