In Vitro Performance Characteristics of High-Frequency Oscillatory Ventilators
- 15 September 2001
- journal article
- research article
- Published by American Thoracic Society in American Journal of Respiratory and Critical Care Medicine
- Vol. 164 (6) , 1019-1024
- https://doi.org/10.1164/ajrccm.164.6.2005008
Abstract
This study aimed to examine the performance characteristics of four high-frequency oscillatory-type ventilators, using an in vitro model of the intubated neonatal respiratory system. Each ventilator was examined across its operative range of settings and at varying model lung compliance (C) and resistance. The oscillatory pressure waveform was measured at the airway opening (Pao). Tidal volume (VT) and flow were determined from pressure changes within the model lung (DeltaPA). The spectral content of the Pao waveform differed between ventilators. The maximum ventilator VT ranged from 3.7 to 11.1 ml at 15 Hz and a mean airway pressure (Paw) of 12 cm H(2)O to oscillate a model lung (C = 0.4 ml/cm H(2)O) through a 3.0-mm internal diameter (i.d.) endotracheal tube (ETT). A small drop in C was associated with a decrease in VT and marked increase in DeltaPA from 0.1 to 0.8 ml/cm H(2)O. The influence of C on VT and DeltaPA and the pressure cost of ventilation (DeltaPA/f.VT(2)) was dependent on the oscillatory frequency, ETT inner diameter, and the specific ventilator used. Substantive differences exist between oscillatory ventilators that need to be considered in their clinical application. The rapid establishment of optimal lung volume and oscillatory frequency is important in minimizing barotrauma during high-frequency oscillatory ventilation.Keywords
This publication has 23 references indexed in Scilit:
- Assessment of high-frequency neonatal ventilator performancesIntensive Care Medicine, 1997
- A multicenter randomized trial of high frequency oscillatory ventilation as compared with conventional mechanical ventilation in preterm infants with respiratory failureEarly Human Development, 1993
- Frequency, Tidal Volume, and Mean Airway Pressure Combinations that Provide Adequate Gas Exchange and Low Alveolar Pressure during High Frequency Oscillatory Ventilation in RabbitsPediatric Research, 1990
- Role of lung volume in lung injury: HFO in the atelectasis-prone lungActa Anaesthesiologica Scandinavica, 1989
- High-Frequency Oscillatory Ventilation in Premature Infants with Respiratory FailureAnesthesia & Analgesia, 1987
- Alveolar Pressure Swings during High Frequency Ventilation in RabbitsPediatric Research, 1985
- Transport of gases in high-frequency ventilationCritical Care Medicine, 1984
- Respiratory system impedance from 4 to 40 Hz in paralyzed intubated infants with respiratory disease.Journal of Clinical Investigation, 1983
- Tidal Volume and Frequency Dependence of Carbon Dioxide Elimination by High-Frequency VentilationNew England Journal of Medicine, 1981
- Effects of frequency, tidal volume, and lung volume on CO2 elimination in dogs by high frequency (2-30 Hz), low tidal volume ventilation.Journal of Clinical Investigation, 1981