Evaluation of the Datascope ACCUSAT pulse oximeter in healthy adults
- 1 January 1988
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 4 (1) , 59-63
- https://doi.org/10.1007/bf01618108
Abstract
Arterial hemoglobin oxygen saturation measured with the Datascope ACCUSAT pulse oximeter was compared with simultaneous arterial hemoglobin oxygen saturation measurements in healthy adult volunteers. One hundred thirty-five arterial blood samples ranging in saturation from 63 to 100% were obtained from 15 adults, aged 20 to 43 years. These subjects had different skin pigmentation, hematocrit, and smoking habits. Steady-state hypoxia was achieved by varying the inspired oxygen concentration between 10 and 21%. Readings from the Datascope ACCUSAT pulse oximeter and the Hewlett-Packard 47201A ear oximeter were compared with arterial blood samples analyzed by the Instrumentation Laboratories IL 282 in vitro CO-Oximeter. The equation for the best fitted linear regression line between the ACCUSAT pulse oximeter and the reference IL 282 CO-Oximeter was: ACCUSAT=1.08(IL)−6.86. The linear regression analysis revealed a high degree of correlation (r=0.99), and a small standard error of the estimate (SEE=1.29%). Simultaneous comparisons between arterial hemoglobin oxygen saturation measured with the ACCUSAT pulse oximeter and the Hewlett-Packard ear oximeter also showed a close correlation (r=0.99, SEE=1.47%). A similar comparison between the ACCUSAT and the Ohmeda 3700 pulse oximeter revealed good correlation (r=0.99, SEE=1.72%). We found that the ACCUSAT pulse oximeter is an accurate instrument for measuring arterial hemoglobin oxygen saturation noninvasively in the range between 60 and 100%.Keywords
This publication has 4 references indexed in Scilit:
- Evaluation of Pulse OximetryAnesthesiology, 1983
- Spectrophotometric monitoring of arterial oxygen saturation in the fingertipMedical & Biological Engineering & Computing, 1980
- ACCURACY, SENSITIVITY TO CARBOXYHEMOGLOBIN, AND SPEED OR RESPONSE OF THE HEWLETT-PACKARD 47201A EAR OXIMETERPublished by Elsevier ,1979
- EAR OXIMETRY - ACCURACY AND PRACTICABILITY IN ASSESSMENT OF ARTERIAL OXYGENATIONPublished by Elsevier ,1976