Experimental and clinical evaluation of a noninvasive reflectance pulse oximeter sensor
- 1 October 1992
- journal article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 8 (4) , 257-266
- https://doi.org/10.1007/bf01617907
Abstract
The objective of this study was to evaluate a new reflectance pulse oximeter sensor. The prototype sensor consists of 8 light-emitting diode (LED) chips (4 at 665 nm and 4 at 820 nm) and a photodiode chip mounted on a single substrate. The 4 LED chips for each wavelength are spaced at 90-degree intervals around the substrate and at an equal radial distance from the photodiode chip. An optical barrier between the photodiode and LED chips prevents a direct coupling effect between them. Near-infrared LEDs (940 nm) in the sensor warm the tissue. The microthermocouple mounted on the sensor surface measures the temperature of the skin-sensor interface and maintains it at a preset level by servoregulating the current in the 940-nm LEDs. An animal study and a clinical study were performed. In the animal study, 5 mongrel dogs (weight, 10–20 kg) were anesthetized, mechanically ventilated, and cannulated. In each animal, arterial oxygen saturation (SaO2) was measured continuously by a standard transmission oximeter probe placed on the dog's earlobe and a reflectance oximeter sensor placed on the dog's tongue. In the first phase of the experiment, signals from the reflectance sensor were recorded while the dog was immersed in ice water until its body temperature decreased to 30°C. In the second phase, the animal's body temperature was normal, and the oxygen content of the ventilator was varied to alter the SaO2. In the clinical study, 18 critically ill patients were monitored perioperatively with the prototype reflectance sensor. The first phase of the study investigated the relationship between local skin temperature and the accuracy of oximeter readings with the reflectance sensor. Each measurement was taken at a high saturation level as a function of local skin temperature. The second phase of the study compared measurements of oxygen saturation by a reflectance oximeter (SpO2[r]) with those made by a co-oximeter (SaO2[IL]) and a standard transmission oximeter (SpO2[t]). Linear regression analysis was used to determine the degree of correlation between (1) the pulse amplitude and skin temperature; (2) SpO2(r) and SaO2(IL); and (3) SpO2(t) and SaO2(IL). Student'st test was used to determine the significance of each correlation. The mean and standard deviation of the differences were also computed. In the animal study, pulse amplitude levels increased concomitantly with skin temperature (at 665 nm,r=0.9424; at 820 nm,r=0.9834;p2(r) correlated well with SaO2(IL) (r=0.982; SEE=2.54%;p<0.001). The results of the clinical study are consistent with these findings. The proto-type reflectance pulse oximeter sensor can yield accurate measurements of oxygen saturation when applied to the forehead or cheek. It is, therefore, an effective alternative to transmission oximeters for perioperative monitoring of critically ill patients.Keywords
This publication has 13 references indexed in Scilit:
- Design and Evaluation of a Reflectance Oxygen Sensor in Critically Ill PatientsPublished by Springer Nature ,1992
- Skin reflectance pulse oximetry: In vivo measurements from the forearm and calfJournal of Clinical Monitoring and Computing, 1991
- In vivo reflectance of blood and tissue as a function of light wavelengthIEEE Transactions on Biomedical Engineering, 1990
- Design and evaluation of a new reflectance pulse oximeter sensor.1988
- Accuracy of Response of Six Pulse Oximeters to Profound HypoxiaAnesthesiology, 1987
- Pulse Oximetry: Technical Aspects of Machine DesignInternational Anesthesiology Clinics, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Noninvasive measurement of beat-to-beat vascular viscoelastic properties in human fingers and forearmsMedical & Biological Engineering & Computing, 1985
- Evaluation of Pulse OximetryAnesthesiology, 1983
- New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger. Part 1: Model experimentMedical & Biological Engineering & Computing, 1982