Estimation of Blood Pressure Variability From 24-Hour Ambulatory Finger Blood Pressure
- 1 July 1998
- journal article
- research article
- Published by Wolters Kluwer Health in Hypertension
- Vol. 32 (1) , 52-58
- https://doi.org/10.1161/01.hyp.32.1.52
Abstract
Abstract —Portapres is a noninvasive, beat-to-beat finger blood pressure (BP) monitor that has been shown to accurately estimate 24-hour intra-arterial BP at normal and high BPs. However, no information is available on the ability of this device to accurately track ambulatory BP variability. In 20 ambulatory normotensive and hypertensive subjects, we measured 24-hour BP by Portapres and through a brachial artery catheter. BP and pulse interval variabilities were quantified by (1) the SDs of the mean values (overall variability) and (2) spectral power, computed either by fast Fourier transform and autoregressive modeling of segments of 120-second duration for spectral components from 0.025 to 0.50 Hz or in a very low frequency range (between 0.00003 and 0.01 Hz) by broadband spectral analysis. The 24-hour SD of systolic BP obtained from Portapres (24±2 mm Hg) was greater than that obtained intra-arterially (17±1 mm Hg, P P P P <0.01). No differences were observed between Portapres and intra-arterial recordings for any estimation of pulse interval variabilities. The overestimation of BP variability by Portapres remained constant over virtually the entire 24-hour recording period. Thus, although clinical studies are still needed to demonstrate the clinical relevance of finger BP variability, our study shows that Portapres can be used with little error to estimate 24-hour BP variabilities if diastolic and mean BPs are used. For systolic BP, the greater error can be minimized by using correction factors.Keywords
This publication has 11 references indexed in Scilit:
- Effects of sino-aortic denervation on spectral characteristics of blood pressure and pulse interval variability: a wide-band approachMedical & Biological Engineering & Computing, 1996
- Prognostic value of 24-hour blood pressure variabilityJournal Of Hypertension, 1993
- Spectral and sequence analysis of finger blood pressure variability. Comparison with analysis of intra-arterial recordings.Hypertension, 1993
- An analysis of the relationship between central aortic and peripheral upper limb pressure waves in manEuropean Heart Journal, 1993
- Feasibility of ambulatory, continuous 24-hour finger arterial pressure recording.Hypertension, 1993
- Sequential spectral analysis of 24-hour blood pressure and pulse interval in humans.Hypertension, 1990
- Comparison of finger and intra-arterial blood pressure monitoring at rest and during laboratory testing.Hypertension, 1989
- Evaluation of the baroreceptor-heart rate reflex by 24-hour intra-arterial blood pressure monitoring in humans.Hypertension, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Continuous vs intermittent blood pressure measurements in estimating 24-hour average blood pressure.Hypertension, 1983