Concurrent solvent recondensation large sample volume splitless injection
- 4 November 2003
- journal article
- research article
- Published by Wiley in Journal of Separation Science
- Vol. 26 (17) , 1491-1498
- https://doi.org/10.1002/jssc.200301553
Abstract
Concurrent Solvent Recondensation Large Sample Volume (CRS‐LV) splitless injection overcomes the limitation of the maximum sample volume to 1–2 μL valid for classical splitless injection. It is based on control of the evaporation rate in the vaporizing chamber, utilization of a strong pressure increase in the injector resulting from solvent evaporation, and greatly accelerated transfer of the sample vapors from the injector into the inlet of an uncoated precolumn by recondensation of the solvent. The sample vapors are transferred into the column as rapidly as they are formed in the injector (concurrent transfer). 20–50 μL of liquid sample is injected with liquid band formation. The sample liquid is received by a small packing of deactivated glass wool positioned slightly above the column entrance at the bottom of the vaporizing chamber. Solvent evaporation strongly increases the pressure in the injector (auto pressure surge), provided the septum purge outlet is closed and the accessible volumes around the vaporizing chamber are small, driving the first vapors into the precolumn. Transfer continues to be fast because of recondensation of the solvent, obtained by keeping the oven temperature below the pressure‐corrected solvent boiling point. The uncoated precolumn must have sufficient capacity to retain most of the sample as a liquid. The experimental data show virtually complete absence of discrimination of volatile or high boiling components as well as high reproducibility.Keywords
This publication has 24 references indexed in Scilit:
- GC instrumentation for on-column injection of large volumes: Automated optimization of conditionsJournal of Microcolumn Separations, 1995
- Inserts providing complete sample evaporation above the column entrance in vaporizing GC injectorsJournal of High Resolution Chromatography, 1993
- Sample evaporation in conventional split/splitless GC injectors. Part 3: Retaining the liquid in the vaporizing chamberJournal of High Resolution Chromatography, 1992
- Improving splitless injection with electronic pressure programmingJournal of High Resolution Chromatography, 1991
- Direct injection of large sample volumes into capillary columns with packed column injectorJournal of High Resolution Chromatography, 1990
- PTV splitless injection of sample volumes up to 20 μlJournal of High Resolution Chromatography, 1988
- Multipurpose cold injector for high resolution gas chromatographyJournal of High Resolution Chromatography, 1987
- On-column injection of large sample volumes usimg the retention gap technique in capillary gas-cromatographyJournal of Chromatography A, 1985
- The influence of the syringe needle on the precision and accuracy of vaporizing GC injectionsJournal of High Resolution Chromatography, 1979
- Splitless injection and the solvent effectJournal of High Resolution Chromatography, 1978