Quality assurance in IMRT: Importance of the transmission through the jaws for an accurate calculation of absolute doses and relative distributions
- 8 February 2002
- journal article
- Published by Wiley in Medical Physics
- Vol. 29 (3) , 269-274
- https://doi.org/10.1118/1.1446110
Abstract
The goal of IMRT is to achieve an isodose distribution conformed to the tumor while avoiding the organs at risk. For these tasks several gantry angles are selected, each one containing a series of different leaf configurations for the multileaf collimator (MLC) (segments). Verifying the relative distributions as well as the absolute doses is an important step for quality assurance issues. We have observed that an accurate modeling of the transmission of the primary x-ray fluence through the jaws and MLC as well as the head scatter is crucial for a precise calculation of relative doses and monitor units. Also, an inaccurate calculation of the output factor for small size segments can lead to important differences in the absolute dose for points under these segments. Incorrect models could lead to systematic errors of around 5% to 10% in the calculated monitor units and a shift in the isodose curves.Keywords
This publication has 9 references indexed in Scilit:
- Calculation of x-ray transmission through a multileaf collimatorMedical Physics, 2000
- Minimizing static intensity modulation delivery time using an intensity solid paradigmInternational Journal of Radiation Oncology*Biology*Physics, 1999
- Verification and implementation of dynamic wedge calculations in a treatment planning system based on a dose-to-energy-fluence formalismMedical Physics, 1996
- Modeling transmission and scatter for photon beam attenuatorsMedical Physics, 1995
- Collimator scatter in photon therapy beamsMedical Physics, 1995
- Realization and verification of three-dimensional conformal radiotherapy with modulated fieldsInternational Journal of Radiation Oncology*Biology*Physics, 1994
- Analytic modeling of photon scatter from flattening filters in photon therapy beamsMedical Physics, 1994
- X-ray field compensation with multileaf collimatorsInternational Journal of Radiation Oncology*Biology*Physics, 1994
- Combining multileaf fields to modulate fluence distributionsInternational Journal of Radiation Oncology*Biology*Physics, 1993