Validation of a simple mechanical accelerometer (pedometer) for the estimation of walking activity
- 1 January 1987
- journal article
- research article
- Published by Springer Nature in European Journal of Applied Physiology
- Vol. 56 (3) , 323-330
- https://doi.org/10.1007/bf00690900
Abstract
A small (28 g) mechanical accelerometer has been tested by subjecting it to controlled bench tests consisting of repetitive vertical oscillations on two designs of test rig. The accelerometer's 3-digit display provided a cumulated score with a maximum of 99.9 units. This score was compared with an independent count of the imposed oscillations and found to be linear with time (r=0.996) and reproducible on retest (coefficient of variation=± 1.5%). The sensitivity ranged from 6.2 to 7.4 units/10,000 oscillations. The response was related to the maximal applied acceleration (calculated from the amplitude and frequency of the oscillations on the assumption that they were sinusoidal) and independent of the amplitude and frequency used. The threshold maximal acceleration was less than 2 m s−2 and the response had reached a plateau at 4 ms−2. During field studies the accelerometer was firmly attached over the hip in a waistband where it responded to the vertical accelerations produced by walking. When compared with an independent count of footsteps from a heel-mounted resistance pad the accelerometer score (after calibration) was not significantly different. The mean difference was (0.29±0.67, S.D.) 103 “steps” in a younger group (n=8, mean age 39 years) and (0.46±1.08, S.D.) 103 “steps” in an older group of women (n=6, mean age 65 years). Scores of around 10×103 “steps” can be expected in a day in moderately active young subjects and 40 × 103 “steps” in a week in the elderly. Simultaneously recorded scores from both right and left hips were not significantly different. In young subjects the mean difference was typically (1.08±0.8) 103 steps in a daily score of 10 × 103 steps and for old subjects (3.5±10.4) 103 steps in a cumulated 6 day score of 40×103 steps. The accelerometer can therefore be used as a pedometer to give an estimate of the number of footsteps taken over long periods.Keywords
This publication has 16 references indexed in Scilit:
- Correlations of pedometer readings with energy expenditure in workers during free-living daily activitiesEuropean Journal of Applied Physiology, 1986
- OBJECTIVE MEASUREMENTS OF CUSTOMARY PHYSICAL ACTIVITY IN ELDERLY MEN AND WOMEN BEFORE AND AFTER RETIREMENTQuarterly Journal of Experimental Physiology, 1986
- Accuracy of Pedometer in Walking and RunningResearch Quarterly for Exercise and Sport, 1980
- Accuracy of Pedometers for Measuring Distance WalkedResearch Quarterly. American Alliance for Health, Physical Education and Recreation, 1977
- The use of pedometer and actometer in studying daily physical activity in man. Part I: Reliability of pedometer and actometerEuropean Journal of Applied Physiology, 1977
- Validity and reliability of pedometers in habitual activity researchEuropean Journal of Applied Physiology, 1977
- Activity Patterns of Elderly Men and WomenJournal of Gerontology, 1977
- A simple load transducer suitable for long-term recording of activity patterns in human subjects.1973
- Active Leisure Time Activities as Related to Age Among Males in a Total PopulationJournal of Gerontology, 1968
- THE ENERGY EXPENDITURE AND MECHANICAL ENERGY DEMAND IN WALKINGErgonomics, 1960