A method for the determination of mercury(II) ions at trace levels is described. The method is based on the profound inhibitory effect of mercury on the enzyme urease. The activity of the enzyme was determined by the rate of ammonia produced from urea as followed by an ammonia gas sensitive iridium thin metalfilm-oxide-semiconductor (IrTMOS) structure. Two systems were investigated. For the initial urease activity studies, a simple microcell was used. Also, a test plate, containing dry reagent strips with all necessary chemicals was developed, making the analytical procedure very simple to perform. The test volume applied was 2 μl and the sensitivity to standards of mercury(II) ions is at least 0.005 μM (1.0 ng/ml). One sample could be analyzed in less than 8 minutes. Furthermore, the kinetics of sensor response versus enzyme activity is discussed.Keywords
This publication has 15 references indexed in Scilit:
- Determination of trace metals in natural waters at nanogram per liter levesls by electrothermal atomic absorption spectrometry after extraction with sodium diethyldithiocarbamateAnalytica Chimica Acta, 1987
- Determination of mercury traces by differential-pulse stripping voltammetry after sorption of mercury vapour on a gold-plated electrodeAnalytica Chimica Acta, 1987
- Decomposition of Marine Biological Tissues for Determination of Arsenic, Selenium, and Mercury Using Hydride-Generation and Cold-Vapor Atomic Absorption SpectrometriesAnalytical Chemistry, 1985
- Enzyme-amplified determination of methotrexate with a pCO2 membrane electrodeAnalytical Chemistry, 1984
- Modified palladium metal-oxide-semiconductor structures with increased ammonia gas sensitivityApplied Physics Letters, 1983
- Dry reagent chemistries in clinical analysisAnalytical Chemistry, 1983
- Enzyme thermistor analysis of heavy metal ions with use of immobilized ureaseFEBS Letters, 1978
- The microbe thermistorNature, 1977
- Preparation of Immobilized Cholinesterase for Use in Analytical Chemistry.Analytical Chemistry, 1965
- Dangerous Properties of Industrial MaterialsSoil Science, 1958