Validation of a hand-held lactate device in determination of blood lactate in critically injured patients
- 1 September 1998
- journal article
- research article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 26 (9) , 1523-1528
- https://doi.org/10.1097/00003246-199809000-00019
Abstract
Admission blood lactate is an accurate predictor of injury severity and mortality in trauma patients. The purpose of this study was to evaluate a portable lactate analyzer in a clinical setting by patient care staff. A prospective, single-operator control solution and patient sample study, using two test devices and a reference device. An urban Level I trauma center. A convenience sample of 47 trauma patients. Intra-assay precision was demonstrated by performance of consecutive analyses of two lactate control solutions (high and low lactate control concentrations) by medical students and physicians. Split sample, simultaneous testing of the portable lactate analyzer was then performed on 66 whole blood specimens from a convenience sample of 47 trauma patients admitted to an urban Level 1 trauma center over 4 mos. Samples were tested simultaneously tested on two portable lactate analyzers and a reference Instrument. Acceptable intra-assay precision was achieved. Regression analysis for two test instruments demonstrated a slope of 0.920, an intercept of 0.323, an r2 of .982, and an SEM of 0.496. Regression analysis for test instrument "A" vs. the reference instrument showed a slope of 0.861, an intercept of 0.209, an r2 of .977, and an SEM of 0.598. Regression analysis for test instrument "B" vs. the reference instrument demonstrated a slope of 0.929, an intercept of -0.095, an r2 of .983, and an SEM of 0.506. Good correlation with a low SEM was obtained over a wide range of clinically relevant lactate values. Use of point of care lactate analysis will decrease analytic time, making an important diagnostic parameter immediately available in the critical care setting. (Crit Care Med 1998; 26:1523-1528)Keywords
This publication has 18 references indexed in Scilit:
- Use of Recombinant Hemoglobin Solution in Reversing Lethal Hemorrhagic Hypovolemic Oxygen Debt ShockThe Journal of Trauma: Injury, Infection, and Critical Care, 1997
- Precision, Accuracy, and Managed Care Implications of a Hand-Held Whole Blood Analyzer in the Prehospital SettingAmerican Journal of Clinical Pathology, 1996
- Bedside measurement of factor VIII:C activity in individuals with hemophilia AAmerican Journal of Hematology, 1996
- Comparing methods of measurement: why plotting difference against standard method is misleadingThe Lancet, 1995
- PROPER HANDLING OF BLOOD SAMPLES FOR LACTATE DETERMINATIONS: A PROSPECTIVE, CONTROLLED TRIAL.Critical Care Medicine, 1995
- Lactic acidosis in critical illnessCritical Care Medicine, 1992
- Early Physiologic Predictors of Injury Severity and Death in Blunt Multiple TraumaArchives of Surgery, 1990
- Rapid measurement of creatine kinase activity in a coronary care unit using a portable benchtop reflectance photometer.BMJ, 1985
- Measurement in Medicine: The Analysis of Method Comparison StudiesJournal of the Royal Statistical Society: Series D (The Statistician), 1983
- Excess Lactate: An Index of Reversibility of Shock in Human PatientsScience, 1964