High Radiation-Flux, Absolute, Water-Flow Calorimeter
- 1 December 1957
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 28 (12) , 1084-1086
- https://doi.org/10.1063/1.1715809
Abstract
A high radiation‐flux, measuring calorimeter has been designed for use in imaging furnaces. It consists of a black‐body receiver provided with a cooled aperture and cylindrical housing. It can be exposed to flux densities of the order of 400 cal/cm2 sec for extended periods without deterioration. Thermopiles measure the temperatures of the inlet and outlet cooling water. The calorimeter is an absolute measuring device with a response time of one minute. The design virtually eliminates heat losses and permits the measurement of a wide range of heat fluxes. The probable error in the radiation‐flux measurement is ±5%.Keywords
This publication has 2 references indexed in Scilit:
- Theoretical Considerations on Performance Characteristics of Solar FurnacesJournal of Jet Propulsion, 1957
- Absolute Water Flow Calorimeter for the Measurement of Intense Beams of Radiant EnergyReview of Scientific Instruments, 1954