Why a Large Tip Electrode Makes a Deeper Radiofrequency Lesion:
- 1 January 1998
- journal article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 9 (1) , 47-54
- https://doi.org/10.1111/j.1540-8167.1998.tb00866.x
Abstract
Increase in RF Lesion Depth with Larger Electrode. Introduction: Increasing electrode size allows an increase in radiofrequency lesion depth. The purpose of this study was to examine the roles of added electrode cooling and electrode-tissue interface area in producing deeper lesions. Methods and Results: In 10 dogs, the thigh muscle was exposed and superfused with heparinized blood. An 8-French catheter with 4- or 8-mm tip electrode was positioned against the muscle with a blood flow of 350 mL/min directed around the electrode. Radiofrequency current was delivered using four methods: (1) electrode perpendicular to the muscle, using variable voltage to maintain the electrode-tissue interface temperature at 60°C; (2) same except the surrounding blood was stationary; (3) perpendicular electrode position, maintaining tissue temperature (3.5-mm depth) at 90°C; and (4) electrode parallel to the muscle, maintaining tissue temperature at 90°C. Electrode-tissue interface temperature, tissue temperature (3.5- and 7.0-mm depths), and lesion size were compared between the 4- and 8-mm electrodes in each method. In Methods 1 and 2, the tissue temperatures and lesion depth were greater with the 8-mm electrode. These differences were smaller without blood flow, suggesting the improved convective cooling of the larger electrode resulted in greater power delivered to the tissue at the same electrode-tissue interface temperature. In Method 3 (same tissue current density), the electrode-tissue interface temperature was significantly lower with the 8-mm electrode. With parallel orientation and same tissue temperature at 3.5-mm depth (Method 4), the tissue temperature at 7.0-mm depth and lesion depth were greater with the 8-mm electrode, suggesting increased conductive heating due to larger volume of resistive heating because of the larger electrode-tissue interface area. Conclusion: With a larger electrode, both increased cooling and increased electrode-tissue interface area increase volume of resistive heating and lesion depth.Keywords
This publication has 22 references indexed in Scilit:
- Thermal Latency in Radiofrequency AblationCirculation, 1996
- Asymptomatic Cardiac Ischemia Pilot (ACIP) StudyCirculation, 1995
- Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With a Saline-Irrigated Electrode Versus Temperature Control in a Canine Thigh Muscle PreparationCirculation, 1995
- Temperature-guided radiofrequency catheter ablation with very large distal electrodes.Circulation, 1993
- Catheter ablation of atrioventricular junction using radiofrequency current in 17 patients. Comparison of standard and large-tip catheter electrodes.Circulation, 1991
- Radiofrequency Catheter Ablation: The Effect of Electrode Size on Lesion Volume In VivoPacing and Clinical Electrophysiology, 1990
- Observations on electrode-tissue interface temperature and effect on electrical impedance during radiofrequency ablation of ventricular myocardium.Circulation, 1990
- Electrode radius predicts lesion radius during radiofrequency energy heating. Validation of a proposed thermodynamic model.Circulation Research, 1990
- Tissue Heating During Radiofrequency Catheter Ablation: A Thermodynamic Model and Observations in Isolated Perfused and Superfused Canine Right Ventricular Free WallPacing and Clinical Electrophysiology, 1989
- Catheter Ablation of Canine Myocardium with Radiofrequency EnergyPacing and Clinical Electrophysiology, 1989