The article gives the original data and analysis of the data of a survey of laboratories in the USA on the precision of the prothrombin time measurement. Criteria for acceptable performance are used to evaluate the performance of laboratories with respect to the reagents used, the technic used, and the size of the hospital. Preliminary recommendations are made as to acceptable performance for commercial thromboplastins.This publication has 0 references indexed in Scilit: