An assessment of eddy current sensitivity and correction in single-shot diffusion-weighted imaging
- 21 November 2000
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 45 (12) , 3821-3832
- https://doi.org/10.1088/0031-9155/45/12/322
Abstract
Artefacts caused by eddy currents are a major problem in diffusion-weighted imaging. This is particularly acute in experiments in which a number of images with differing degrees of diffusion weighting and/or differently oriented diffusion-weighting gradients need to be combined. The echo-planar imaging sequence is particularly sensitive to the effects of residual eddy currents, especially due to the low bandwidth in the phase-encoding direction. Two published schemes are investigated regarding the effectiveness of eddy current correction. That of Jezzard et al (1998 Magn. Reson. Med. 39 801-12) requires the acquisition of additional experimental data in order to perform a post-acquisition correction, whereas that of Wider et al (1994 J. Magn. Reson. A 108 255-8) attempts to reduce the eddy currents directly. It is found that the latter experiment gives a somewhat superior performance and a combination of the two approaches results in an almost complete elimination of artefact. An alternative single-shot imaging experiment to echo-planar imaging is given by sequences based on fast spin-echo methods, which should be insensitive to the effects of constant eddy currents. It is shown that the intrinsic eddy-current-related artefact level in such experiments is indeed low, residual artefacts being attributed to eddy current decay during the echo train. In situations of poor main magnetic field homogeneity or large eddy currents such sequences may be gainfully used instead of echo-planar imaging.Keywords
This publication has 32 references indexed in Scilit:
- Correction of eddy current-induced artefacts in diffusion tensor imaging using iterative cross-correlationMagnetic Resonance Imaging, 1999
- A simplified method to measure the diffusion tensor from seven MR imagesMagnetic Resonance in Medicine, 1998
- MRI white matter diffusion anisotropy and PET metabolic rate in schizophreniaNeuroReport, 1998
- Elimination of eddy current artifacts in diffusion‐weighted echo‐planar images: The use of bipolar gradientsMagnetic Resonance in Medicine, 1997
- Correction for distortion of echo‐planar images used to calculate the apparent diffusion coefficientMagnetic Resonance in Medicine, 1996
- Anisotropy of water diffusion in the myocardium of the rat.Circulation Research, 1994
- Estimation of the Effective Self-Diffusion Tensor from the NMR Spin EchoJournal of Magnetic Resonance, Series B, 1994
- MR diffusion tensor spectroscopy and imagingBiophysical Journal, 1994
- A PFG NMR experiment for accurate diffusion and flow studies in the presence of eddy currentsJournal of Magnetic Resonance (1969), 1991
- Multiecho imaging sequences with low refocusing flip anglesJournal of Magnetic Resonance (1969), 1988