Diffusion-based model of pulse oximetry: in vitro and in vivo comparisons
- 1 March 1994
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 33 (7) , 1279-1285
- https://doi.org/10.1364/ao.33.001279
Abstract
A model of pulse oximetry is developed based on the three-dimensional photon diffusion theory. To test the applicability of the model, an in vitro assay was developed. Three different scattering levels and six different relative dye concentrations were analyzed. Average percent errors of 13.9% were obtained over the full range of the study. An in vivo clinical study of two pulse oximeter probes with different spectral characteristics was compared with results estimated by the model. The model correctly predicted the changes in pulse oximeter response resulting from the wavelength changes. A χ2 test gave a probability of 20% that the model fit the data. These results demonstrated the utility of the photon diffusion theory for the modeling of tissue optics.Keywords
This publication has 7 references indexed in Scilit:
- Simple photon diffusion analysis of the effects of multiple scattering on pulse oximetryIEEE Transactions on Biomedical Engineering, 1991
- Errors in 14 pulse oximeters during profound hypoxiaJournal of Clinical Monitoring and Computing, 1989
- Comparison of time-resolved and -unresolved measurements of deoxyhemoglobin in brain.Proceedings of the National Academy of Sciences, 1988
- Effects of multiple scattering and peripheral circulation on arterial oxygen saturation measured with a pulse-type oximeterMedical & Biological Engineering & Computing, 1984
- Experimental test of the reduced effective velocity of light in a diffuse mediumOptics Letters, 1980
- Spectrophotometric monitoring of arterial oxygen saturation in the fingertipMedical & Biological Engineering & Computing, 1980
- Generalized standard addition methodAnalytical Chemistry, 1979