Performance of QRS detection for cardiac magnetic resonance imaging with a novel vectorcardiographic triggering method
Open Access
- 19 October 2000
- journal article
- clinical trial
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 12 (5) , 678-688
- https://doi.org/10.1002/1522-2586(200011)12:5<678::aid-jmri4>3.0.co;2-5
Abstract
In many cardiac patients, image quality and/or scan efficiency is reduced due to imprecise R‐wave ability to trigger the scan due to noise on the electrocardiogram (ECG) caused by the magnetic resonance (MR) environment. We developed a triggering system that uses the spatial information of the vectorcardiogram (VCG) to minimize the effects of MR‐related noise on triggering. Fifteen volunteers underwent standard cardiovascular MR exams, and a total of 52,474 R‐waves were evaluated with the algorithm, giving a performance index of 99.91%. The mean propagation delay of the system was −10.64 ± 3.19 msec, which falls within the real‐time definition for cardiac MRI triggering. Five patients had arrhythmias consisting of premature ventricular depolarizations (PVDs) and supraventricular extra systoles. For those patients with PVDs, all arrhythmic beats were rejected unless they passed through the algorithm's reference point. The performance index for the arrhythmic patients approached 100%. VCG‐based triggering has been demonstrated to provide near 100% triggering performance during cardiac MR examinations. J. Magn. Reson. Imaging 2000;12:678–688.Keywords
This publication has 11 references indexed in Scilit:
- Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitionsMagnetic Resonance in Medicine, 1999
- Improved myocardial tagging contrastMagnetic Resonance in Medicine, 1993
- An approach to QRS complex detection using mathematical morphologyIEEE Transactions on Biomedical Engineering, 1993
- A comparison of the noise sensitivity of nine QRS detection algorithmsIEEE Transactions on Biomedical Engineering, 1990
- Suppression of radiofrequency interference in cardiac gated MRI: A simple designMagnetic Resonance in Medicine, 1988
- Monitoring of acutely III patients during nuclear magnetic resonance imaging: Use of a Time‐varying filter electrocardiographic gating device to reduce gradient artifactsMagnetic Resonance in Medicine, 1988
- Cardiovascular alterations in Macaca monkeys exposed to stationary magnetic fields: Experimental observations and theoretical analysisBioelectromagnetics, 1983
- The effect of rapid rise-time magnetic fields on the ECG of the ratClinical Physics and Physiological Measurement, 1982
- Observation of blood flow E.M.F. in externally applied strong magnetic field by surface electrodesMedical & Biological Engineering & Computing, 1967
- An Accurate, Clinically Practical System For Spatial VectorcardiographyCirculation, 1956