Accurate calibration of gold film standards by neutron activation and gravimetry
- 1 December 1978
- journal article
- Published by AIP Publishing in Journal of Applied Physics
- Vol. 49 (12) , 5903-5908
- https://doi.org/10.1063/1.324556
Abstract
A set of evaporated gold films on silicon substrates has been prepared for use as standards in the measurement of film thickness by Rutherford backscattering. The mass of gold in the films was determined independently by gravimetry and by instrumental neutron activation. The area was measured with an optical comparator; film uniformity was assayed by backscattering. Systematic errors in the single-element comparator method of activation analysis have been assessed and either eliminated or compensated, with the result that the single-sample agreement in mass between the two methods was 0.3% and the mean relative bias was below the random errors, also 0.3%.This publication has 11 references indexed in Scilit:
- Radial efficiency gradients in Ge(Li) gamma detectorsJournal of Radioanalytical and Nuclear Chemistry, 1977
- Off center gamma ray detection efficiencies of cylindrical single open ended (SOE) Ge(Li) detectorsNuclear Instruments and Methods, 1975
- Evaluation of lyophilization for the preconcentration of natural water samples prior to neutron activation analysisAnalytical Chemistry, 1975
- Characterization of a nuclear reactor for neutron activation analysisJournal of Radioanalytical and Nuclear Chemistry, 1974
- Experimental comparison of multi-channel analyzer. Dead-time corrections for short-lived radionuclidesJournal of Radioanalytical and Nuclear Chemistry, 1974
- The accurate determination of Ga in Gax Sey by means of neutron activation: A comparison with X-ray fluorescence and wet chemical methodsJournal of Radioanalytical and Nuclear Chemistry, 1974
- A new circuit for a dead time compensating timerNuclear Instruments and Methods, 1972
- The limit of precision in nuclear and analytical chemistryNuclear Instruments and Methods, 1972
- Coincidence losses in activation analysisJournal of Radioanalytical and Nuclear Chemistry, 1971
- Computation of lithium-drifted germanium detector peak areas for activation analysis and .gamma.-ray spectrometryAnalytical Chemistry, 1968